FLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051

FDA recall event 91953 · Class II · Richard Wolf GmbH

Key facts

Event ID 91953
Recall numbers Z-1422-2023, Z-1423-2023, Z-1424-2023, Z-1425-2023
Recalling firm Richard Wolf GmbH
Product type Devices
Products FLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051, FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 828.051, FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 8735.685, FLEX. GRASP. FORCEPS 6.6FR WL 550MM, Product Number 8736.685
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions…
Initiation date Mar 24, 2023
Report date May 3, 2023
City Knittlingen
Country Germany