Drug recall

Ultraview SL (UVSL) Command Module, Model 91496

FDA recall event 91851 · Class II · Spacelabs Healthcare, Inc.

Key facts

Event ID 91851
Recall numbers Z-1503-2023
Recalling firm Spacelabs Healthcare, Inc.
Product type Devices
Products Ultraview SL (UVSL) Command Module, Model 91496
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have…
Initiation date Apr 12, 2023
Report date May 10, 2023
City Snoqualmie
State WA
Country United States
FEI number 3010157426

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)