Drug recall

Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system…

FDA recall event 91782 · Class II · Moberg Research, Inc.

Key facts

Event ID 91782
Recall numbers Z-1338-2023
Recalling firm Moberg Research, Inc.
Product type Devices
Products Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system that collects, displays, and stores multiple physiological measurements Model: CNS-350
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin. DOC
Initiation date Feb 20, 2023
Report date Apr 12, 2023
City Ambler
State PA
Country United States
FEI number 3007880955

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls