Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801…

FDA recall event 91602 · Class I · Datascope Corp.

Key facts

Event ID 91602
Recall numbers Z-1145-2023, Z-1146-2023
Recalling firm Datascope Corp.
Product type Devices
Products Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34…, Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Classification Class I
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication…
Initiation date Feb 7, 2023
Report date Mar 15, 2023
City Mahwah
State NJ
Country United States
FEI number 3001418283