Drug recall
ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815…
FDA recall event 91182 · Class II · Siemens Medical Solutions USA, Inc
Key facts
| Event ID | 91182 |
|---|---|
| Recall numbers | Z-0472-2023 |
| Recalling firm | Siemens Medical Solutions USA, Inc |
| Product type | Devices |
| Products | ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815 Intended Use: Family of linear accelerator systems is to deliver X-Ray photon and electron… |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | When selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter |
| Initiation date | Nov 9, 2022 |
| Report date | Dec 21, 2022 |
| City | Malvern |
| State | PA |
| Country | United States |
| FEI number | 2240869 |