Drug recall

Hobbs Medical Polypectomy Snare, Catalog No. 7202

FDA recall event 90456 · Class II · Hobbs Medical, Inc.

Key facts

Event ID 90456
Recall numbers Z-1388-2022
Recalling firm Hobbs Medical, Inc.
Product type Devices
Products Hobbs Medical Polypectomy Snare, Catalog No. 7202
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall During use, the sheath of the device may accordion at the proximal end where the sheath meets the shrink band. Kinking of the sheath may prevent the snare from fully closing, and injury may result.
Initiation date Jun 3, 2022
Report date Jul 27, 2022
City Stafford Springs
State CT
Country United States
FEI number 1220592

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls