Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of…
FDA recall event 90433 · Class II · Breas Medical, Inc.
Key facts
| Event ID | 90433 |
|---|---|
| Recall numbers | Z-1382-2022 |
| Recalling firm | Breas Medical, Inc. |
| Product type | Devices |
| Products | Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation Model Number: 230000 ¿ If the recalled product… |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health… |
| Initiation date | Jun 2, 2022 |
| Report date | Jul 20, 2022 |
| City | North Billerica |
| State | MA |
| Country | United States |
| FEI number | 3010817335 |