Drug recall
Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000…
FDA recall event 90368 · Class II · GE Healthcare, LLC
Key facts
| Event ID | 90368 |
|---|---|
| Recall numbers | Z-1511-2022 |
| Recalling firm | GE Healthcare, LLC |
| Product type | Devices |
| Products | Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA… |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table… |
| Initiation date | Jun 2, 2022 |
| Report date | Aug 17, 2022 |
| City | Waukesha |
| State | WI |
| Country | United States |
| FEI number | 2126677 |