RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14…
FDA recall event 89800 · Class II · RAYSEARCH LABORATORIES AB
Key facts
| Event ID | 89800 |
|---|---|
| Recall numbers | Z-0849-2022 |
| Recalling firm | RAYSEARCH LABORATORIES AB |
| Product type | Devices |
| Products | RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.7.6.7, 4.9.0.42, 5.0.1.11, 5.0.2.35… |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Support or Fixation may be unexpected.The density in… |
| Initiation date | Sep 22, 2021 |
| Report date | Apr 6, 2022 |
| City | Stockholm |
| Country | Sweden |
| FEI number | 3007774465 |