Drug recall

Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION…

FDA recall event 89784 · Class II · B Braun Medical Inc

Key facts

Event ID 89784
Recall numbers Z-0872-2022, Z-0873-2022
Recalling firm B Braun Medical Inc
Product type Devices
Products Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION 2, Article Code Number 3456102A. These conductivity sensors may be used as spare parts for…, Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be used as spare parts for…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or…
Initiation date Jan 28, 2022
Report date Apr 20, 2022
Termination date May 15, 2024
City Bethlehem
State PA
Country United States