Drug recall

BenchMark ULTRA and DISCOVERY ULTRA Instruments

FDA recall event 89727 · Class II · Ventana Medical Systems Inc

Key facts

Event ID 89727
Recall numbers Z-0894-2022
Recalling firm Ventana Medical Systems Inc
Product type Devices
Products BenchMark ULTRA and DISCOVERY ULTRA Instruments
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.
Initiation date Jan 20, 2022
Report date Apr 13, 2022
City Oro Valley
State AZ
Country United States
FEI number 1000125183

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: