Drug recall

TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDNS090Z

FDA recall event 89170 · Class II · Acclarent, Inc.

Key facts

Event ID 89170
Recall numbers Z-0415-2022
Recalling firm Acclarent, Inc.
Product type Devices
Products TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDNS090Z
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Suction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect…
Initiation date Nov 12, 2021
Report date Dec 29, 2021
City Irvine
State CA
Country United States
FEI number 3005172759

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls