Drug recall

Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063…

FDA recall event 88854 · Class II · Philips North America Llc

Key facts

Event ID 88854
Recall numbers Z-0213-2022
Recalling firm Philips North America Llc
Product type Devices
Products Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray…
Initiation date Sep 24, 2021
Report date Nov 17, 2021
Termination date Sep 26, 2024
City Cambridge
State MA
Country United States
FEI number 3016618143

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls