Drug recall

EVIS EXERA BF-XT160 Bronchofibervideoscope, Model No. BF-XT160

FDA recall event 88560 · Class II · Olympus Corporation of the Americas

Key facts

Event ID 88560
Recall numbers Z-2497-2021
Recalling firm Olympus Corporation of the Americas
Product type Devices
Products EVIS EXERA BF-XT160 Bronchofibervideoscope, Model No. BF-XT160
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall A gluing step was not performed during the manufacturing of the affected device's insertion tube. There is a potential for the insertion tube to develop a leak, which poses an infection control risk.
Initiation date Aug 17, 2021
Report date Sep 29, 2021
Termination date Jan 12, 2024
City Center Valley
State PA
Country United States
FEI number 2429304

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: