Drug recall

CombiDiagnost R90 Software Version R1.0 and R1.1

FDA recall event 88393 · Class II · Philips Medical Systems Gmbh, DMC

Key facts

Event ID 88393
Recall numbers Z-2357-2021
Recalling firm Philips Medical Systems Gmbh, DMC
Product type Devices
Products CombiDiagnost R90 Software Version R1.0 and R1.1
Classification Class II
Status Ongoing
Voluntary or mandated FDA Requested
Reason for recall Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This…
Initiation date Jul 22, 2020
Report date Sep 8, 2021
City Hamburg
Country Germany
FEI number 3003768251

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls