Incisive CT Computed Tomography X-ray System Model: 728143

FDA recall event 87427 · Class II · Philips North America Llc

Key facts

Event ID 87427
Recall numbers Z-1265-2021
Recalling firm Philips North America Llc
Product type Devices
Products Incisive CT Computed Tomography X-ray System Model: 728143
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward…
Initiation date Feb 22, 2021
Report date Mar 24, 2021
Termination date Feb 9, 2023
City Cambridge
State MA
Country United States
FEI number 3016618143

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)