Drug recall

DJO EMPOWR KNEE Femoral Sizer

FDA recall event 87332 · Class II · Encore Medical, LP

Key facts

Event ID 87332
Recall numbers Z-1524-2021
Recalling firm Encore Medical, LP
Product type Devices
Products DJO EMPOWR KNEE Femoral Sizer
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Incorrect assembly of a subcomponent by a vendor and as a result, the windows provided to show the size are facing the opposite direction.
Initiation date Feb 5, 2021
Report date May 12, 2021
City Austin
State TX
Country United States
FEI number 1000116912

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: