Drug recall
BARD MARQUEE Disposable Core Biopsy Instruments XX.Xcm, XX.Xcm, XXg x XXcm REF XXXXXXX Note - for definition…
FDA recall event 87316 · Class II · Bard Peripheral Vascular Inc
Key facts
| Event ID | 87316 |
|---|---|
| Recall numbers | Z-1289-2021 |
| Recalling firm | Bard Peripheral Vascular Inc |
| Product type | Devices |
| Products | BARD MARQUEE Disposable Core Biopsy Instruments XX.Xcm, XX.Xcm, XXg x XXcm REF XXXXXXX Note - for definition of the "X's refer to Tyvek Lid Label.pdf" uploaded in Associated Documents |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use. |
| Initiation date | Jan 14, 2021 |
| Report date | Mar 31, 2021 |
| Termination date | Jan 28, 2025 |
| City | Tempe |
| State | AZ |
| Country | United States |
| FEI number | 2020394 |