Drug recall
BioFire BCID2 Panel; IVD, Rx Only Part No: RFIT-ASY-0147 / UDI: 00815381020338 - Product Usage: intended for…
FDA recall event 87282 · Class II · BioFire Diagnostics, LLC
Key facts
| Event ID | 87282 |
|---|---|
| Recall numbers | Z-1157-2021, Z-1158-2021 |
| Recalling firm | BioFire Diagnostics, LLC |
| Product type | Devices |
| Products | BioFire BCID2 Panel; IVD, Rx Only Part No: RFIT-ASY-0147 / UDI: 00815381020338 - Product Usage: intended for use with FilmArray systems., Gram-Negative Bacteria And Associated Resistance Markers; Film Array Blood Culture Identification (BCID); Film Array BCID Panel - Product Usage: intended for use with FilmArray systems. Part No… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | The firm has identified an increased risk of false positive Pseudomonas aeruginosa when BCID/BCID2 Panels are used with specific BD BACTEC" blood culture vials. |
| Initiation date | Jan 26, 2021 |
| Report date | Mar 10, 2021 |
| Termination date | May 2, 2022 |
| City | Salt Lake City |
| State | UT |
| Country | United States |
| FEI number | 3002773840 |