Drug recall

Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or…

FDA recall event 87153 · Class III · Biomeme, Inc.

Key facts

Event ID 87153
Recall numbers Z-0940-2021
Recalling firm Biomeme, Inc.
Product type Devices
Products Nucleic acid-based in vitro diagnostic devices for the detection of 2019-novel coronavirus (2019-nCoV) and/or diagnosis of 2019-nCoV infection in human clinical specimens. The RT-PCR test runs on a…
Classification Class III
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.
Initiation date Jan 5, 2021
Report date Feb 10, 2021
Termination date Jun 11, 2021
City Philadelphia
State PA
Country United States
FEI number 3014117298

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.