Drug recall

Contour VL Variable Length Ureteral Stent Set, UPN M006180155090, M006180156090, M006180156180, M006180156190

FDA recall event 86880 · Class II · Boston Scientific Corporation

Key facts

Event ID 86880
Recall numbers Z-0757-2021, Z-0758-2021
Recalling firm Boston Scientific Corporation
Product type Devices
Products Contour VL Variable Length Ureteral Stent Set, UPN M006180155090, M006180156090, M006180156180, M006180156190, Contour VL Variable Length Ureteral Stent, UPN M0061801550, M0061801560, M0061801570
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall Trace amounts of metallic debris were found on the stent related to the manufacturing process. The debris was confirmed to be particles of nickel.
Initiation date Nov 16, 2020
Report date Jan 6, 2021
City Marlborough
State MA
Country United States
FEI number 3005099803

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls