Drug recall

Olympus Medical Systems Corporation (OMSC) Evis Exera II Bronchovideoscope Olympus BF Type Q180 (BF-Q180) All…

FDA recall event 86384 · Class II · Olympus Corporation of the Americas

Key facts

Event ID 86384
Recall numbers Z-3017-2020
Recalling firm Olympus Corporation of the Americas
Product type Devices
Products Olympus Medical Systems Corporation (OMSC) Evis Exera II Bronchovideoscope Olympus BF Type Q180 (BF-Q180) All serial numbers - Product Usage: designed to be used with an Olympus video system center…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The BF-Q180 bronchoscope is being recalled because it does not have a 510(k) clearance, and the device is associated with a higher rate of patient infections than other comparable OMSC bronchoscopes.
Initiation date Aug 31, 2020
Report date Oct 7, 2020
Termination date Feb 14, 2024
City Center Valley
State PA
Country United States
FEI number 2429304

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: