Drug recall

AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter…

FDA recall event 86282 · Class II · ESAOTE S.P.A

Key facts

Event ID 86282
Recall numbers Z-3052-2020, Z-3053-2020
Recalling firm ESAOTE S.P.A
Product type Devices
Products AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is…, AC/DC POWER SUPPLY for 7400, Part Number 151004301, XP Power Model Number AHM180PS19-XE1057, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical…
Initiation date Sep 8, 2020
Report date Oct 7, 2020
Termination date Apr 17, 2023
City Genova
Country Italy
FEI number 3014563045

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls