Drug recall

ProxiDiagnost N90, UDI 706100

FDA recall event 85603 · Class II · Philips North America, LLC

Key facts

Event ID 85603
Recall numbers Z-2256-2020
Recalling firm Philips North America, LLC
Product type Devices
Products ProxiDiagnost N90, UDI 706100
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats…
Initiation date Apr 23, 2020
Report date Jun 10, 2020
Termination date Mar 30, 2023
City Andover
State MA
Country United States
FEI number 1218950

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls