Drug recall
Accelerator a3600 Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The…
FDA recall event 85469 · Class II · Inpeco S.A.
Key facts
| Event ID | 85469 |
|---|---|
| Recall numbers | Z-2198-2020, Z-2199-2020, Z-2200-2020 |
| Recalling firm | Inpeco S.A. |
| Product type | Devices |
| Products | Accelerator a3600 Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation., Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation., FlexLab Augomation System using the Aliquoter Module (Inpeco P/N FLX-212) The systems consolidate multiple Analytical instruments into a unified workstation. |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume…, Potential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is… |
| Initiation date | Apr 7, 2020 |
| Report date | Jun 10, 2020 |
| City | Lugano |
| Country | Switzerland |
| FEI number | 3010825766 |
At a glance
- 1–9 Products in this recall
- None FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.