Drug recall

Accelerator a3600 Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The…

FDA recall event 85469 · Class II · Inpeco S.A.

Key facts

Event ID 85469
Recall numbers Z-2198-2020, Z-2199-2020, Z-2200-2020
Recalling firm Inpeco S.A.
Product type Devices
Products Accelerator a3600 Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation., Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation., FlexLab Augomation System using the Aliquoter Module (Inpeco P/N FLX-212) The systems consolidate multiple Analytical instruments into a unified workstation.
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume…, Potential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is…
Initiation date Apr 7, 2020
Report date Jun 10, 2020
City Lugano
Country Switzerland
FEI number 3010825766

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls