Drug recall

Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ Aspiration System, UDI…

FDA recall event 85424 · Class II · INARI MEDICAL

Key facts

Event ID 85424
Recall numbers Z-2299-2020
Recalling firm INARI MEDICAL
Product type Devices
Products Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ Aspiration System, UDI: 00850291007079
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The hemostasis valve of aspiration guide catheter devices may fail to seal resulting in a loss of vacuum during thrombus aspiration, which may result the procedure being prolonged as another device…
Initiation date Mar 23, 2020
Report date Jun 17, 2020
Termination date Aug 13, 2021
City Irvine
State CA
Country United States
FEI number 3011525976

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: