Drug recall

iChem¿VELOCITY Urine Chemistry System-

FDA recall event 85423 · Class II · Beckman Coulter Inc.

Key facts

Event ID 85423
Recall numbers Z-1944-2020, Z-1945-2020
Recalling firm Beckman Coulter Inc.
Product type Devices
Products iChem¿VELOCITY Urine Chemistry System-, iQ200 Series Urine Microscopy Analyzer- All part numbers
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA…, A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA…
Initiation date Mar 30, 2020
Report date May 20, 2020
Termination date Aug 31, 2023
City Brea
State CA
Country United States
FEI number 2050012

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: