Drug recall

0.035"(.89mm)x150 cm guidewire. Model 114135150. Soft Tip 3.5cm. REF/UDI: 114135150/(01)20886333008074…

FDA recall event 85398 · Class I · Argon Medical Devices, Inc

Key facts

Event ID 85398
Recall numbers Z-1791-2020, Z-1792-2020
Recalling firm Argon Medical Devices, Inc
Product type Devices
Products 0.035"(.89mm)x150 cm guidewire. Model 114135150. Soft Tip 3.5cm. REF/UDI: 114135150/(01)20886333008074. RxOnly.STERILE EO, 0.035"(.89mm)x80 cm guidewire. Model 114135080. Soft Tip 3.5cm. REF/UDI: 114135080/(01)208863333008067. RxOnly STERILE EO
Classification Class I
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The stiffness of the soft end of the guidewire caused tissue perforation.
Initiation date Dec 4, 2019
Report date May 13, 2020
Termination date Apr 1, 2024
City Athens
State TX
Country United States
FEI number 1625425

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls