Drug recall

Varian Multileaf Collimator [MLC]; Millenium Multileaf Collimator [52, 80, 120 leaf]; High Definition 120…

FDA recall event 85269 · Class II · Varian Medical Systems, Inc.

Key facts

Event ID 85269
Recall numbers Z-1899-2020
Recalling firm Varian Medical Systems, Inc.
Product type Devices
Products Varian Multileaf Collimator [MLC]; Millenium Multileaf Collimator [52, 80, 120 leaf]; High Definition 120 Multileaf Collimator Model Number(s): product code: H54 and H56; MLC software version 8.5…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall After a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move during an arc treatment. The issue is related to a transient…
Initiation date Aug 12, 2019
Report date May 13, 2020
Termination date Jan 21, 2021
City Palo Alto
State CA
Country United States
FEI number 2916710

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: