Drug recall

natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph

FDA recall event 85189 · Class II · Natus Neurology DBA Excel Tech., Ltd. (XLTEK)

Key facts

Event ID 85189
Recall numbers Z-1630-2020
Recalling firm Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
Product type Devices
Products natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the…
Initiation date Mar 5, 2020
Report date Apr 8, 2020
Termination date Aug 9, 2021
City Oakville
Country Canada
FEI number 3002809369

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls