Drug recall

ARTISTE MV System Model #8139789 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13

FDA recall event 85161 · Class II · Siemens Medical Solutions USA, Inc

Key facts

Event ID 85161
Recall numbers Z-1677-2020, Z-1678-2020, Z-1679-2020, Z-1680-2020
Recalling firm Siemens Medical Solutions USA, Inc
Product type Devices
Products ARTISTE MV System Model #8139789 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13, MEVATRON M2/Primus Mid-Energy Model #01940035, MEVATRON PRIMART #05500371, MEVATRON KD-2 #08515520, MEVATRON KD2 #01940753, MEVATRON M-2 6740 #09401506, MEVATRON MD-2 #09401654, MEVATRON MDX-2…, ONCOR Avante Garde #5863472, ONCOR Expression #7360717, ONCOR Impression #5857920, ONCOR Impression Plus #5857912 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13, PRIMUS HI #04504200 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
Initiation date Feb 18, 2020
Report date Apr 15, 2020
Termination date Feb 9, 2021
City Malvern
State PA
Country United States
FEI number 2240869

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls