Drug recall

Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00…

FDA recall event 85090 · Class II · Philips North America, LLC

Key facts

Event ID 85090
Recall numbers Z-1616-2020
Recalling firm Philips North America, LLC
Product type Devices
Products Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2…
Initiation date Feb 20, 2020
Report date Apr 8, 2020
Termination date Sep 21, 2021
City Andover
State MA
Country United States
FEI number 1218950

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls