Drug recall
QIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling and…
FDA recall event 85004 · Class II · Qiagen Sciences LLC
Key facts
| Event ID | 85004 |
|---|---|
| Recall numbers | Z-1476-2020 |
| Recalling firm | Qiagen Sciences LLC |
| Product type | Devices |
| Products | QIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling and real-time detection of PCR assays. |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false… |
| Initiation date | Jan 28, 2020 |
| Report date | Mar 18, 2020 |
| Termination date | Sep 26, 2024 |
| City | Germantown |
| State | MD |
| Country | United States |
| FEI number | 3003572099 |