Drug recall

QIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling and…

FDA recall event 85004 · Class II · Qiagen Sciences LLC

Key facts

Event ID 85004
Recall numbers Z-1476-2020
Recalling firm Qiagen Sciences LLC
Product type Devices
Products QIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling and real-time detection of PCR assays.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false…
Initiation date Jan 28, 2020
Report date Mar 18, 2020
Termination date Sep 26, 2024
City Germantown
State MD
Country United States
FEI number 3003572099

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls