Drug recall

Iliac Screws, multiple sizes, Article Nos. B36020500 B36020570 B36020580 B36020590 B36027500 B36027570…

FDA recall event 84928 · Class II · Medicrea International

Key facts

Event ID 84928
Recall numbers Z-1472-2020, Z-1473-2020
Recalling firm Medicrea International
Product type Devices
Products Iliac Screws, multiple sizes, Article Nos. B36020500 B36020570 B36020580 B36020590 B36027500 B36027570 B36027580 B36027590 B36028500 B36028570 B36028580 B36028590 B36029500 B36029570 B36029580…, Pedicle Screws, multiple sizes, Article Nos. B36004525 B36004530 B36004535 B36004540 B36004545 B36004550 B36004555 B36005530 B36005535 B36005540 B36005545 B36005550 B36005555 B36005560 B36006530…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.
Initiation date Jan 28, 2020
Report date Mar 18, 2020
Termination date Mar 17, 2023
City Rillieux La Pape
Country France
FEI number 1000432246

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls