Drug recall

Rheumatoid Factor (RF) Kit for use on SPAPLUS, for in-vitro diagnostic use, Product code: LK151.S.A., UDI…

FDA recall event 84909 · Class II · The Binding Site Group, Ltd.

Key facts

Event ID 84909
Recall numbers Z-1500-2020
Recalling firm The Binding Site Group, Ltd.
Product type Devices
Products Rheumatoid Factor (RF) Kit for use on SPAPLUS, for in-vitro diagnostic use, Product code: LK151.S.A. , UDI: 05051700018319
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall It has been identified that the rheumatoid factor (RF) calibrator values over recover in comparison to the reference material RF serum. This could result in reporting falsely abnormal results which…
Initiation date Nov 6, 2019
Report date Mar 25, 2020
Termination date Sep 8, 2022
City Birmingham
Country United Kingdom
FEI number 3002808340

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: