Drug recall

UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The…

FDA recall event 84186 · Class II · Beckman Coulter, Inc.

Key facts

Event ID 84186
Recall numbers Z-0516-2020
Recalling firm Beckman Coulter, Inc.
Product type Devices
Products UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results…
Initiation date Oct 7, 2019
Report date Dec 4, 2019
Termination date Apr 27, 2021
City Miami
State FL
Country United States
FEI number 3000203293

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls