Drug recall

HeartStart XL+ Defibrillator/Monitor, Model 861290

FDA recall event 84115 · Class II · Philips North America, LLC

Key facts

Event ID 84115
Recall numbers Z-0498-2020
Recalling firm Philips North America, LLC
Product type Devices
Products HeartStart XL+ Defibrillator/Monitor, Model 861290
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The rotary therapy selector switch may fail, resulting in unexpected device behavior. These behaviors include: The device may not turn on; The device may not perform the selected function; The device…
Initiation date Oct 16, 2019
Report date Nov 27, 2019
Termination date Nov 29, 2023
City Andover
State MA
Country United States
FEI number 1218950

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.