Drug recall

LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for…

FDA recall event 84059 · Class II · Smith & Nephew, Inc.

Key facts

Event ID 84059
Recall numbers Z-0351-2020
Recalling firm Smith & Nephew, Inc.
Product type Devices
Products LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully…
Initiation date Oct 7, 2019
Report date Nov 20, 2019
Termination date Mar 8, 2023
City Memphis
State TN
Country United States
FEI number 1020279

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls