Drug recall

Epix Electrosurgical Probes with Smoke Evacuation, Angled L-Hook Tip, 5mm x 42 cm REF CW002 QTY: 1 - Product…

FDA recall event 83909 · Class II · Applied Medical Resources Corp

Key facts

Event ID 83909
Recall numbers Z-0509-2020, Z-0510-2020
Recalling firm Applied Medical Resources Corp
Product type Devices
Products Epix Electrosurgical Probes with Smoke Evacuation, Angled L-Hook Tip, 5mm x 42 cm REF CW002 QTY: 1 - Product Usage: The Epix Electrosurgical Probe with Smoke Evacuation is intended to deliver energy…, GelPOINT Path Transanal Procedural Pack with Epix Electrosurgical L-Hook with Smoke Evacuation, 4x5.5cm and 5mm x 42cm REF TA211 Product Usage: The GelPOINT Path Transanal Access Platform is indicated…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Unintentional shifting of the tip insulation of the firm's electrosurgical probes.
Initiation date Dec 12, 2018
Report date Nov 27, 2019
Termination date Jan 22, 2021
City Rancho Santa Margarita
State CA
Country United States
FEI number 2027111

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls