Drug recall

SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system

FDA recall event 83883 · Class II · Siemens Medical Solutions USA, Inc

Key facts

Event ID 83883
Recall numbers Z-0265-2020
Recalling firm Siemens Medical Solutions USA, Inc
Product type Devices
Products SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional…
Initiation date Sep 24, 2019
Report date Nov 27, 2019
Termination date Jul 8, 2024
City Malvern
State PA
Country United States
FEI number 2240869

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: