Drug recall

Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis…

FDA recall event 83774 · Class II · Centurion Medical Products Corporation

Key facts

Event ID 83774
Recall numbers Z-0120-2020, Z-0121-2020
Recalling firm Centurion Medical Products Corporation
Product type Devices
Products Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 310CR, Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 330CR
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.
Initiation date Aug 28, 2019
Report date Oct 16, 2019
Termination date Jul 23, 2020
City Williamston
State MI
Country United States
FEI number 3008403546

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls