Drug recall
IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of…
FDA recall event 83358 · Class II · Medical Intelligence Medizintechnik Gmbh
Key facts
| Event ID | 83358 |
|---|---|
| Recall numbers | Z-2082-2019 |
| Recalling firm | Medical Intelligence Medizintechnik Gmbh |
| Product type | Devices |
| Products | IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan. |
| Initiation date | Jul 18, 2019 |
| Report date | Jul 31, 2019 |
| Termination date | Aug 11, 2023 |
| City | Schwabmunchen |
| Country | Germany |
| FEI number | 3003251135 |