Drug recall

IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of…

FDA recall event 83358 · Class II · Medical Intelligence Medizintechnik Gmbh

Key facts

Event ID 83358
Recall numbers Z-2082-2019
Recalling firm Medical Intelligence Medizintechnik Gmbh
Product type Devices
Products IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.
Initiation date Jul 18, 2019
Report date Jul 31, 2019
Termination date Aug 11, 2023
City Schwabmunchen
Country Germany
FEI number 3003251135

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: