Drug recall

GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a…

FDA recall event 83342 · Class II · Xcision Medical Systems, LLC

Key facts

Event ID 83342
Recall numbers Z-2081-2019
Recalling firm Xcision Medical Systems, LLC
Product type Devices
Products GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.
Initiation date Jul 3, 2019
Report date Jul 31, 2019
Termination date Jul 1, 2021
City Columbia
State MD
Country United States
FEI number 3014273653

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls