Drug recall

Spacelabs Healthcare Smart Disclosure System, Model 92810, a component of the Intesys Clinical Suite (ICS) G2.

FDA recall event 83220 · Class II · Spacelabs Healthcare, Inc.

Key facts

Event ID 83220
Recall numbers Z-2035-2019
Recalling firm Spacelabs Healthcare, Inc.
Product type Devices
Products Spacelabs Healthcare Smart Disclosure System, Model 92810, a component of the Intesys Clinical Suite (ICS) G2.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Several reports were received that patient records were printed with the correct patient demographics but containing another patient's waveforms.
Initiation date Jan 28, 2019
Report date Jul 31, 2019
Termination date Feb 24, 2022
City Snoqualmie
State WA
Country United States
FEI number 3010157426

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls