Drug recall
Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse Control Panel Arm, Model. No. MAJ-1664
FDA recall event 83171 · Class II · Olympus Corporation of the Americas
Key facts
| Event ID | 83171 |
|---|---|
| Recall numbers | Z-1983-2019 |
| Recalling firm | Olympus Corporation of the Americas |
| Product type | Devices |
| Products | Olympus WM-NP2 or WM-DP2 Mobile Workstations with Nurse Control Panel Arm, Model. No. MAJ-1664 |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | WM-NP/DP2 workstations could become unstable and may fall over if the control panel arm is mounted with a video monitor weighing more than 19.8 pounds. |
| Initiation date | Jun 10, 2019 |
| Report date | Jul 24, 2019 |
| Termination date | Jul 10, 2020 |
| City | Center Valley |
| State | PA |
| Country | United States |
| FEI number | 2429304 |
At a glance
- 1–9 Products in this recall
- 10–99 FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.
Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: