Drug recall

Cook Kwart Retro-Inject Ureteral Stent Set: PRODUCT/CATALOG NUMBER/GPN: 4.7 Fr 003500 G14916; 6.0 Fr 003600…

FDA recall event 82795 · Class II · Cook Inc.

Key facts

Event ID 82795
Recall numbers Z-1369-2019, Z-1370-2019
Recalling firm Cook Inc.
Product type Devices
Products Cook Kwart Retro-Inject Ureteral Stent Set: PRODUCT/CATALOG NUMBER/GPN: 4.7 Fr 003500 G14916; 6.0 Fr 003600 G14836; 7.0 Fr 003700 G14837; 8.0 Fr 003800 G14844; Set with AQ¿ Hydrophilic Coating…, Cook Salle Intraoperative Pyeloplasty Stent Set PRODUCT/CATALOG NUMBER/GPN: 4.0 Fr - SIPSF-040018-56-6 G18168 4.0 Fr - SIPSF-040018-59 G32773 4.7 Fr SIPSF-050018-59 G32774
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Kwart RetroInject Stent Set (Multi-Length) and Salle Intraoperative Pyeloplasty Stent Set Instructions for Use (IFU) did not include a warning for stent knotting.
Initiation date Apr 22, 2019
Report date May 29, 2019
Termination date May 5, 2020
City Bloomington
State IN
Country United States
FEI number 1820334

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls