Drug recall

Philips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A, M3536M, M3536MC…

FDA recall event 82750 · Class II · Philips North America, LLC

Key facts

Event ID 82750
Recall numbers Z-1923-2019
Recalling firm Philips North America, LLC
Product type Devices
Products Philips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring…
Initiation date Jun 7, 2019
Report date Jul 17, 2019
Termination date May 8, 2020
City Andover
State MA
Country United States
FEI number 1218950

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: