Drug recall

IntelliVue MX40 Patient Monitor, Product number 857146, Service Numbers All: 453564711531 and 453564711541…

FDA recall event 82696 · Class II · Philips North America, LLC

Key facts

Event ID 82696
Recall numbers Z-1692-2019, Z-1693-2019, Z-1694-2019, Z-1695-2019
Recalling firm Philips North America, LLC
Product type Devices
Products IntelliVue MX40 Patient Monitor, Product number 857146, Service Numbers All: 453564711531 and 453564711541 - Product Usage: Used primarily as a traditional telemetry medical device. It connects to the…, IntelliVue MX40 Patient Monitor, Product number 865350, Service # USA: 453564262491 and 453564262511 - Product Usage: Used primarily as a traditional telemetry medical device. It connects to the…, IntelliVue MX40 Patient Monitor, Product number 865351, Service Numbers All other countries: 453564451791, 453564451811, 453564467721, 453564467741, 453564467761, 453564467781, 453564467801…, IntelliVue MX40 Patient Monitor, Product number 865352, Service Numbers All: 453564262571, 453564262591, 453564615311, 453564615331, 453564615351, and 453564615371 Product Usage: Used primarily as a…
Classification Class II
Status Ongoing
Voluntary or mandated Firm Initiated
Reason for recall The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .
Initiation date Mar 28, 2019
Report date Jun 12, 2019
City Andover
State MA
Country United States
FEI number 1218950

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls