Accelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage: The…

FDA recall event 82216 · Class II · Accelerate Diagnostics Inc

Key facts

Event ID 82216
Recall numbers Z-1065-2019
Recalling firm Accelerate Diagnostics Inc
Product type Devices
Products Accelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage: The Accelerate Pheno system is intended to measure signal intensity of fluorescent probes bound…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to…
Initiation date Mar 13, 2017
Report date Apr 3, 2019
Termination date May 21, 2021
City Tucson
State AZ
Country United States
FEI number 3010671651