Drug recall

e500 Automatic Transport Ventilator, Model Number 01EVE500

FDA recall event 82178 · Class I · O-Two Medical Technologies, Inc.

Key facts

Event ID 82178
Recall numbers Z-0976-2019, Z-0977-2019, Z-0978-2019
Recalling firm O-Two Medical Technologies, Inc.
Product type Devices
Products e500 Automatic Transport Ventilator, Model Number 01EVE500, e600 Automatic Transport Ventilator, Model Number 01EVE600, e700 Automatic Transport Ventilator, Model Number 01EVE700
Classification Class I
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it…
Initiation date Feb 12, 2019
Report date Apr 10, 2019
Termination date Jun 29, 2023
City Brampton
Country Canada
FEI number 3002807836

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls